PM-Series

PM-35U

Standard Range 2V

3 Optional DC Input Ranges

from 200mV to 200V

3 1/2 DIGIT with 0.4 LED

A Low Cost Easy To Use General Purpose Meter

General Features

The PM-35U Digital Panel Meter is an economical, high performance instrument incorporating a number of fea tures usually found only on more expensive meters. Utilizing the dual slope method of integration, the unit measures differential and single-ended DC voltages over five user selectable ranges from 199.9mV to 199.9V full scale.

Maximum res o lu tion is 100V over 1999 counts. Provision is made for user con nec tions to provide various operating modes, including a ra ti om et ric voltmeter, current meter, ratiometric ohmmeter, and a tem pera ture difference meter.

The true differential input capability of the Model PM-35U is particularly useful for making accurate measurements of very small signals in the presence of much larger common mode signals.

Because of its high noise immunity, it is also ideal for measuring various balanced transducers and bridge inputs. When measuring bridge circuits, long term drift of the excitation voltage can be compensated for by using the ratiometric voltmeter mode of operation.

Typical Application Connections

SINGLE-ENDED METER200mV Range, >2V Range 200mV Range:

1) Change R6 from 470k to 47k, 5%; 2) Change R9 from 18.2k to 26.7k, 1%; 3) Remove R8 (9.53k) and short with a jumper; 4) It is recom mend ed that C4, C5 and C7 be changed to 0.22mF; 5) Adjust R10 until CV1=100mV.

>2V Range: Install R1 and R2 as specified under section titled Useful Tables.

View more application connections and connection instructions on page 3.

Specifications

Input Configuration:......True differential and single-ended

Full Scale Ranges:........199.9mVDC

1.999VDC (standard)

19.99VDC

199.9VDC

Input Impedance:...........Exceeds 1000M on 200mV and

2V ranges; 10M on all other ranges

Input Protection:............250VDC or 175VAC maximum on

200mV and 2V ranges; 1200VDC or 850VAC on all other ranges

Accuracy:.......................(0.1% of reading + 1 digit)

Temperature Coefficient:..5PPM/C in ratiometric operation; 80

to 100PPM/C using internal reference on 200mV and 2V ranges

Warm Up Time:..............2 minutes to specified accuracy

Conversion Rate:...........3 readings per second nominal

Display:...........................0.4" LED

Decimal Selection:........User programmable to 3 positions

Overrange Indication:...When input exceeds full scale on

any range being used, most significant 1 digit & polarity symbol are displayed with all other digits blank

Power Requirements:...+4.5 to +5.5VDC at 200mA max.

Operating Temperature:....0 to +60C

Storage Temperature:...-20 to +70C

Relative Humidity..........95% (non-condensing)

Case Dimensions:.........Bezel 2.76 x 1.17 (69.75 x 29.7mm)

Depth behind Bezel 3.32(84mm) plus 0.68 (17.27mm) for connector.

Weight:.............................88 gms (3.1 oz)

PM-Series, high performance versatility for a wide range of applications

PM-35A.................... 3.5 digit Red LED, Precision Preference, 2VDC, 5VDC Power PM-35U.................... 3.5 digit Red LED, Low Cost, 2VDC, 5VDC Power

PM-45L................ 4.5 digit Red LED, Precision Meter w/Differential Input PM-45X............... 4.5 digit LCD, Precision Meter w/Differential Input PM-45XU.................3.5 digit LCD, Low Cost Meter w/Differential Input

Page 1

Functional Diagram

Connector Pinouts

The Texmate Model PM-35U inter con nects by means of a standard PC board edge connector having two rows of 10 pins, spaced on 0.156" centers. Connector are available from Texmate, or from almost any connector manu fac turer. Pro vi sion has also been made for direct solder termination to the PC board, thus elimi nat ing the need for a connector. Please note that the warranty will be void if there is any damage due to soldering.

Calibration Pot

Pins A, 1 and 3Decimal Select: Decimal points may be displayed as required by connecting appropriate pin to Decimal Select Common Pins B or 2.

Pins B and 2Decimal Select Common: Decimal points are displayed as required by connecting Pins B or 2 to the appropriate Decimal Select Pins A, 1, or 3.

Pin DReference Input: Reference voltage input for A to D convert er. Normally supplied from Reference Voltage Output Pin 4 which is internally jumpered to Pin D by way of Junction J1. Pin D may be used as an input for ratiometric measurements. Minimum usable voltage is.05VDC with +4.0VDC being the maximum.

Pin EAnalog Common: 2.8V output referenced to +5V System Power Input L, approximately +2.2V output referenced to Power Ground Input Pin K.

Pins F and 6Signal High Input: Signal high input of A to D converter. Maximum overvoltage protection is 250VDC or 175VAC.

Pins H and 7Voltage Divider Signal High Input: Signal high input for voltages that require attenuation or scaling. Dividing resistors R1 and R2 may be mounted internally for voltages up to 200V

Page 2

max. Matched dividing resistors for 2V (1/10) and 200V (1/100) ranges are available from Texmate. Shunt resistors for current measurements up to 200mA may be internally mounted in the R2 position. The current loop is then applied to Signal High Input Pin F and returned through Pin J or Pin 8.

Pins J and 8Signal Low Input: Signal low input of A to D convert er. Maximum overvoltage protection is 250VDC or 175VAC.

Pins K and 9Power Ground Input: Negative terminal of + 5VDC power should be connected to Pin 9. All digital signals, decimal points, display test, and overrange should be returned to this ground point.

Pins L and 10- +5VDC System Power Input: Meter requires regulated (10%) 5VDC power at approximately 200mA.

Pin 4Reference Voltage Output: Internal precision voltage ref er ence. Standard output is 1.000V, adjustable to 5% by R10 potentiometer. Usable voltages from.05V to 2.8V for special high impedance scaling can be obtained by changing the value of internal dividing resistors R8 and R9.

Pin 5Display Test: All segments will light up when Pin 5 is connected to +5VDC System Power Input Pin L or Pin 10.

Component Layout

COMPONENT SIDE

SOLDER SIDE

Signal Conditioning Components

SPAN Potentiometer (Pot)

The SPAN pot is on the right side of the display. Typical adjustment is 20% of the input signal range.

Calibration Procedure

Apply power to the meter. Then with a precision DC reference source, apply +1.900VDC be tween the Sig nal High Input Pin F and the Signal Low Input Pin J. Adjust R10 po ten ti om e ter (on left side as viewed from rear) until the display reads +1.9900V. Note: The voltage ap plied in this case is for a 1.999V F.S. meter. For other ranges, the voltage applied should be similarly pro por tion ate to the particular full scale voltage.

Typical Application Connections

The PM-35U may be used in a wide variety of configurations.

The following circuits illustrate some of the possibilities and demonstrate the exceptional versatility of Texmate products.

Components called for in the applications which are not part of the standard meter may be supplied by the user or in some cases purchased from Texmate. The circuit diagrams explain the basic pinout connections required for each application.

Unless otherwise specified, the diagrams will show the component values and solder junctions that would normally be installed on a standard 2V range meter. For those applications which have alternative ranges and/or input configurations, the required component values and any modifications are

described in the text.

NOTE: Use of these application circuits is entirely at the risk and responsibility of the user and any user modification of the meter may at the discretion of Texmate, void the warranty. (See rear page for user's responsibility and warranty details) The following legend applies to all application circuits: 1) optional component positions are shown in dotted lines; 2) internal solder junctions are shown by for a closed junction or for an open junction; 3) calibration voltages as measured by an external user supplied voltmeter are shown by

Useful Tables

R1 F.S. In

200mV omit

omit 2V

9M 20V

10M 200V

R2 Resolution

omit

omit

100k

100V

1mV

10mV

100mV

Rs External

0.01

0.1

Rs Internal

100

Resolution

10mA

1mA

Resolution

100A

10A

1A

100nA

Page 3

Case Dimensions

PM Case Dimensions and Panel Cutouts

PANEL CUTOUT

FRONT VIEW

SIDE VIEW

TOP VIEW

When extra panel mounting tightness is required, optional Screw Mounting Clips can be purchased seperately and attach to the sliding mounting side clips

option metal

screw mounting clip

Ordering Information

Standard Options for this Model Number

Special Options and Accessories

Part Number Description

4BASIC MODEL NUMBER

PM-35U..... 3.5 digit Red LED, Low Cost, 2VDC, 5VDC pwr

Note: For operation, the PM-35U requires a CN-L10 Dual Row 10 Pin PCB edge connector 0.156 center or equivalent (see accessories).

Part Number Description

4SPECIAL OPTIONS (Specify Inputs & Req. Reading) ZS..........Custom display scaling within standard ranges ZR-200V....200 VDC Range Change.

ZR-20V......20 VDC Range Change.

ZRS-200MV..200 mVDC range change.

ZRS-PMRP...Non-standard range and scale.

4ACCESSORIES

CN-L10...... Dual Row 10 Pin Connector, Solder Type

Page 4